UBT-112 High Frequency Mobile X-ray C-arm System
Specifications:
Fluoroscopic Capacity Max rated capacity : Tube Current 4mA, Tube Voltage 110kV
Automatic Fluoroscopy:
Tube Voltage:40kV~110kV adjust automatically ,Tube Current : 0.3mA~4mA set manually
Manual Fluoroscopy :
Continuous Tube Voltage:40kV~110kV, Continuous Tube Current : 0.3mA~4mA
Pulse Fluoroscopy :
Continuous Tube Voltage 40kV~110kV ,Continuous Tube Current 4.1mA~8mA
Photography Capacity
Max rated capacity : 3.5 KW
Tube Voltage and Current Combination:
40kV~49kV 1 ~125 mAs
50kV~59kV 1 ~110 mAs
60kV~69kV 1 ~90 mAs
70kV~79kV 1 ~80 mAs
80kV~89kV 1 ~71 mAs
90kV~99kV 1 ~63 mAs
100kV~110kV
1 ~40 mAs
Plateholder Size : 200mm×250mm(8″×10″)or 250mm×300mm(10×12)
X-ray Tube Special for High Frequency
Fixed anode
Dual-focus : large focus: 1.5mm , small focus: 0.6mm
Inverter Frequency: 40KHz
Thermal capacity: 35KJ (47KHU)
Video System
Image Intensifier: Image Intensifier made by TOSHIBA (9)
CCD Vidicon : Ultra low-light CCD camera from Japan
Monitor : Horizontal 1000 lines and vertical 800 lines,
Bandwidth : 12.5MHz, Image/sec : 75
CCU(central control) : Recursive Filter: K=8, 8 images storage, image upright, image overturn,positive & negative image;LIH(last image freeze, and OSD(monitor display)
Structure :
Directive wheel: ±90°revolution, can freely change the moving direction of the unit.
Ascending & Descend- ing Range of Pillar: ≥400mm
C-arm:
Forward and Backward Movement: 200mm
Revolution around Horizontal Axis: ±180°
Revolution around Vertical Axis : ±15°
Slip on orbit: 120° (+90°~ -30°)
Packing Details: 2440mm×1100mm×1420mm
N.W. 300kgs G.W.400kgs