S6100D anesthesia system used for patients whom need inhalation anesthesia, controlled breathing, or help to breathe as well as monitoring and display of the patient's ventilation parameters in medical department. Suitable for adult and children above
Gas source:O2(for medical use)
N2O(for medical use)
Ventilation mode :IPPV、A/C、SIMV、SIGH、MANU.
Alarm and protection:
The AC power outage alarm
No tidal volume
High (low) ventilation alarm
High (low)Airway pressure alarm
Pressure limit
Continuous pressure alarm
The maximum limited pressure <12.5 kPa
With O2 drive insufficient pressure alarm and N2O truncation protection
O2, N2O flowmeter has the linkage to ensure O2 concentration≥25%。When O2 closed ,N2O closes at the same time.
Power supply 100-240V~
Frequency 50 /60HZ
Input power 40 VA